Conductivity measurement in demanding water applications needs stable values, even when media are conductive, contaminated or difficult to handle. CONDIX4213 measures the specific conductivity of liquid media and provides process values, identification data and diagnostic data to a master system via RS485 with Modbus RTU protocol.
The 4-electrode measuring principle with a C0.5 1/cm cell constant supports applications up to 500 mS/cm and reduces the influence of pollution, polarization effects and wire resistance. This makes the converter well suited for landfill seepage water, seawater and black water treatment on ships.
Commissioning stays practical: parameters and input configuration are adjustable via the interface or SENSware, while selectable temperature compensation functions adapt the measurement to the application. IP67 protection and PVC-U process material support reliable integration.
Measures specific conductivity of liquid media and records temperature with an integrated Pt1000 measuring element.
Provides conductivity, temperature, system status, parameters, identification data and diagnostics via RS485 Half-Duplex.
Device parameters and input configuration can be adjusted via the interface, SENSware configuration software or MULTICON-H.
Selectable modes include no compensation, linear coefficient, natural waters, ASTM-D1125 ultra-pure water, NaCl diluted solution, ASTM-D5391 acidic pure water and ASTM-D5391 alkaline pure water.
Uses a PVC fitting with cap nut and supports installation with PVC-U standard fittings.
Uses an 8 pole round connector plug with brass nickel plated design
LEDs indicate supply voltage, bus communication and addressing status.
Addressing can be realized with a field attachable female connector or in a junction box.
| Media temperature |
0...60 °C
|
|---|---|
| Pressure rating |
16 bar max.
|
| Cell constant |
0.5 1/cm exact cell constant labeled on the type plate
|
|---|---|
| Conductivity range |
0...500 mS/cm
|
| Accuracy temperature |
0.2 K
|
| Temperature compensation |
Selectable: no temp comp; linear coeff; natural water comp; ASTM-D1125 ultra-pure water; NaCl dilute solution; ASTM-D5391 acidic & alkaline pure water
|
| Temperature range |
-50...+200 °C
|
| Temperature sensor |
Pt1000, Class A
|
| Accuracy Conductivity |
±1 % OR (>2% of measure for <20μS/cm)
|
| Condensation |
not permitted
|
|---|---|
| Degree of protection case |
IP67
|
| Operating temperature |
-10...+60 °C
|
| Storage temperature |
-10...+60 °C
|
| Weight |
160 g
|
|---|
| CE Conformity |
Yes
|
|---|
| Baud rate |
1200, 2400, 4800, 9600, 19200
|
|---|---|
| Interface type |
RS485
|
| Type of communication protocol |
Modbus RTU
|
| Electrical connection type |
8 pole round connector plug M12x1
For 8-pin round plug connector M12x1 IP67: M12 8-pole circular connector |
|---|---|
| Material |
Brass nickel-plated
For 8-pin round plug connector M12x1 IP67: Brass nickel-plated |
| Power supply |
4.7...28 V
|
|---|---|
| Current consumption |
max. 60 mA
|
| Process connection |
PVC fitting with cap nut
|
|---|---|
| Wetted parts |
PVC-U, Casting resin, Graphite
|