CO2 calibration accessory for IAQ monitor zero calibration

HD21AB17.9

CO2 calibration accessory for IAQ monitor calibration, used between the instrument and a nitrogen cylinder for 0 ppm CO2 calibration. Connecting tube included.

Introduction

CO2 calibration work is easier to control when the gas path is guided directly to the instrument. This accessory connects the IAQ monitor with a nitrogen cylinder for CO2 calibration, supporting the 0 ppm calibration procedure described for the HD21ABE17.
 
The accessory slides onto the top of the instrument, with the T/RH probe passing through the central hole. The tube from the nitrogen cylinder connects to the CO2 input of the accessory, helping service personnel carry out the calibration with the correct setup.
 
For the documented CO2 zero calibration procedure, the nitrogen cylinder flow meter is adjusted to a constant flow from 0.3 to 0.5 l/min. The connecting tube is included, reducing the need to source the essential connection separately.

Controlled setup
Connects the instrument to a nitrogen cylinder for documented CO2 calibration.
0 ppm calibration
Supports CO2 zero calibration with nitrogen during the HD21ABE17 procedure.
Direct instrument fit
Slides onto the instrument top while the T/RH probe passes through the central hole.
Tube included
Supplied with the connecting tube for the nitrogen cylinder connection.
Defined gas flow
Procedure specifies a constant nitrogen flow from 0.3 to 0.5 l/min.

Features

CO2 calibration connection

The accessory creates the connection between the IAQ instrument and a nitrogen cylinder for CO2 calibration.

Instrument mounting

The accessory is positioned on the top of the instrument, with the T/RH probe passing through the central hole.

Nitrogen tube connection

The tube from the nitrogen cylinder connects to the CO2 input of the accessory during calibration.

Procedure support

The accessory is used in the documented 0 ppm CO2 calibration process with nitrogen.