Aggressive media such as hydrofluoric acid require an electrode built with the right glass from the start. This combined pH electrode uses special membrane glass that allows use with HF and is specified for scrubber and process chemistry applications.
The electrode measures 0..14 pH and uses an Ag/AgCl reference system with 3 mol/l KCL-Gel and 1 x ceramic diaphragm. Operating limits for HF depend on pH value and F-concentration: 300 ppm at pH 2, 1000 ppm at pH 3, 6000 ppm at pH 4 and no limitation from pH ≥5 at 20°C.
For integration, the electrode provides a PG13,5 process connection, S8 industrial electrical connection and a 240 mm shaft length according to the ordering code. It operates at -5..+70 °C, up to 6 bar and with media conductivity from 150 µS/cm.
| Pressure rating |
6 bar max.
|
|---|
| pH range |
0...14 pH
|
|---|
| Minimum conductivity |
> 150 µS/cm
|
|---|
| Process connection |
PG 13.5
|
|---|
| Electrical connection type |
For S8 Industrial screw plug head: S8
|
|---|
| Probe Diameter |
For 240 mm: 12 mm
|
|---|---|
| Probe length |
For 240 mm: 240 mm
|