Portable conductivity meter for water analysis and temperature

HD2106.x

Portable conductivity meter for water analysis, measuring conductivity, resistivity, TDS, salinity and temperature (Pt100/Pt1000). IP66 housing, automatic calibration and optional data logging up to 36,000 samples.

Introduction

Reliable water quality checks need a meter that is easy to carry, clear to read and ready for changing measurement tasks. This portable conductivity meter supports conductivity, liquid resistivity, TDS, salinity and temperature measurement with combined 4-ring or 2-ring conductivity/temperature probes.
 
Use it in field work, laboratories or water-related process checks where stable, documented readings matter. Automatic calibration with 147 µS/cm, 1413 µS/cm, 12880 µS/cm and 111800 µS/cm standard solutions reduces setup effort, while selectable temperature compensation from 0.00 to 4.00%/°C helps align readings to the sample behavior.
 
The IP66-protected ABS and rubber housing is built for mobile use. Both models transfer real-time measurements via RS232C, and the HD2106.2 adds USB 2.0 connection plus memory for up to 36,000 conductivity and temperature samples.

Multi-parameter use
Covers conductivity, resistivity, TDS, salinity and temperature in one portable meter.
Easy calibration
Automatic calibration uses 147 µS/cm to 111800 µS/cm standard solutions.
Field-ready housing
IP66 protection with ABS and rubber housing supports mobile measurement work.
Logged data
HD2106.2 stores up to 36,000 conductivity and temperature samples.

Features

Probe compatibility

The instruments work with combined 4-ring and 2-ring conductivity/temperature probes, as well as Pt100 or Pt1000 probes for temperature-only measurement.

Temperature compensation

Conductivity readings can be compensated automatically or manually from 0…100 °C, with selectable αT from 0.00 to 4.00%/°C and reference temperature of 20 °C or 25 °C.

Measurement functions

The meter calculates maximum, minimum and average values and includes relative measurement, Auto-HOLD and automatic turn-off, which can also be disabled.

Data communication

Both models include an electrically isolated RS232C serial port for real-time transfer to a PC or portable printer. HD2106.2 also provides USB 2.0 communication.

Data logging

HD2106.2 records up to 2,000 pages with 18 samples each. The storage interval is selectable from 1 second to 1 hour.

Display and operation

The LCD shows the selected conductivity-related parameter and temperature together, with 2x4½ digits plus symbols and a visible area of 52x42 mm.